After establishing a calibration on an x-ray fluorescence
instrument everybody likes to use it as long as possible. This wouldn’t be
any
problem
if the instrument would not show any changes
in sensitivity
over time.
But the opposite is the case. The x-ray tube looses
intensity over time.
The detector looses sensitivity over time too. Service
works like change
of an x-ray tube or a crystal might have an influence
to the sensitivity as well.
The loss on intensity is not constant for all elements.
Light elements from
Be to Zn normally show a higher influence than heavy
elements.
To use an existing calibration over a long time the
change in sensitivity for
an element has to be detected
by a drift monitor and then used to correct
for it.
Such samples are called drift monitors, monitor samples,
reference samples or monitor standards.
A drift monitor must show spezial properties. Such a
sample should not
show any change of element composition over a long time.
Today mostly samples from glass often made from silicate
glass are used. However also monitors made from metals
are used.
FLUXANA
calibrations, installed by us in many customer laboratories
contain a special selection of drift monitors to guarantee
an excellent
long term stability.
These monitor samples were designed especially for the
use in our
customer applications.
Driven by customers we produce monitor samples in our
glass production designed on the customer need.
|


|
| Examples
for FLUXANA monitor glasses: |
| FLX-C1,
C3, Z1, Z2, Z3 |
Cement,
raw materials cement industry, glass industry |
| FLX-S13-F, S5, S7, S8, S9, Z4 |
Multielement
X-ray calibration |
| FLX-SP1, SP1-32, SP-2, SP2-32, D1 |
Raw materials. Ores |
| FLX-S6M, Q1, Q2, Q3, R1, R3, R4, S2 |
X-ray calibration |

Link
to request
Top
|
 |
 |
 |
 |
|